Comments
Rheodyne
front-loading injectors have a built-in needle port with a patented direct-connection
design. It connects the tip of the syringe needle directly to the end
of the sample loop. There is no connecting passage that traps sample,
which can enter the loop when the next sample is loaded. Syringe accuracy
is maintained and flushing after every injection is rarely necessary.
Size designations
of sample loops are nominal. Actual volumes can differ from the nominal
designations because of the tolerance of the tubing bore. Accuracy of
large loops (2 mL) is about 5%, intermediate loops (20 µL) 10%,
and small loops (5 µL) 30%.
The following
Rheodyne Technical Notes contain useful information:
Technical
Note 5, Achieving Accuracy and Precision with Rheodyne Sample Injectors.
Technical
Note 7, Pressure Drop of Valves and Tubes from 1 to 1000 mL/min.
Technical
Note 9, How to Make Tube Connections Between Injectors, Columns
and Detectors in Liquid Chromatography.
|
What
Happens Inside
Mobile phase in the loop is displaced when you dispense sample from
the syringe during loading. The boundary between the sample and mobile
phase has a parabolic profile due to laminar flow; the velocity at the
center of the tubing is twice the average velocity, and at the wall
it is almost zero. So, sample in the center travels 2 µL along the loop
for every 1 µL dispensed.
This behavior
accounts for the shape of the curve in the graph, a plot of sample volume
dispensed from the syringe vs. mass of sample injected into the column.
The curve has three regions.
1. When
the volume dispensed is less than half the loop volume, the curve is
linear. Sample has not yet reached the far end of the loop. Within this
region, performance depends on syringe accuracy and the operator skill.
Accuracy is typically 1%, if you use the syringe full scale volume.
Precision is 0.2% to 2% RSD, depending on your skill. This is the region
of the partial-filling method.
2. When
the volume dispensed is between half a loop volume and about two loop
volumes, the curve is nonlinear. Sample is lost from the loop, so accuracy
and precision are poor. If you dispense a volume equal to the volume
of the loop, you are in this region of inferior performance.
3. When
the volume dispensed is several loop volumes, the mass injected is independent
of the volume dispensed. The loop contains only pure sample, undiluted
by residual mobile phase. Accuracy is determined by the loop (see Tips
Page). Precision is 0.05% to 1%, depending on the volume loaded.
Five loop volumes produce about 0.1% precision. When only a few loop
volumes are loaded, in order to conserve sample, precision is improved
by loading nearly the same volume each time. This is the region of the
complete-filling method.
|