Comments
To
inject a blank: Inject a volume of mobile phase in the same way that you
inject sample. This mobile phase should have exactly the same composition
as that pumped at the time of injection. If your sample is dissolved in
a solvent different from this, then it is instructive to make an additional
blank injection using the sample solvent. This will show if the artifacts
are caused by the difference in solvents.
Spurious
peaks are more likely when operating at high detector sensitivity and
can be a difficult problem to troubleshoot. In some cases they are not
real substances, but artificial signals from the detector. In other cases
they are real substances, but they come from outside of the injector,
either by being unintentionally loaded with the blank, or by being desorbed
from the column due to the injection of the blank. In rare cases they
are substances which have contaminated the internal parts of the injector.
Use only
genuine Rheodyne replacement parts.
|
Artifact
Peaks
Artifacts
are peaks that appear unexpectedly. The problems are separated into
three groups: carryover peaks, spurious peaks, and distorted peaks:
- Carryover
peaks are those produced because of cross contamination from a previous
injection (see Fig. 12). When a blank is injected
a miniature chromatogram appears.
- Spurious
peaks are those that are unrelated to the previously injected sample
(see Fig. 13). When a blank is injected these "ghost"
peaks appear as peaks with retention times that are different from sample
components.
- Distorted
peaks are those that look unfamiliar. A blank injection looks OK, but
when the sample is injected some of the peaks are fronting, tailing,
or split (double). (See Fig. 14.)
Its
best that you read all parts of Artifact Peaks, including
the comments, before taking action. You may also want to consider the
problem is caused by a component other than the injector (see Appendices
C and D).
Fig.
12. Peaks in blank are CARRYOVER from previous sample.
|