Comments
The
UV absorbance of dissolved oxygen itself is very small, but it is believed
that dissolved oxygen forms complexes with solvent molecules, and that
these have appreciable absorbance. With mixed solvents, more than one
complex species would be expected. Indeed, with mixed mobile phases more
than one peak is often observed.
You are more
likely to encounter artifact peaks when using an efficient degassing system.
Oxygen in the air will dissolve in degassed mobile phase very rapidity.
The more thorough the degassing, the greater is the driving force to reabsorb
oxygen. If peaks appear when they have previously been absent, check to
see if a procedure has changed that might have changed the oxygen concentration
in either the mobile phase or the sample. Oxygen can enter the mobile
phase by diffusing upstream against the outflowing helium in the reservoir
vent tubes and by diffusing through Teflon® reservoir-to-pump transfer
tubes.
|
Oxygen
Peak Artifacts
Artifact peaks are sometimes produced when the sample (or blank) contains
oxygen in a concentration that differs from the oxygen concentration
in the mobile phase that is flowing at the time of injection. This difference
is common when the mobile phase is completely degassed, such as when
continuously sparged with helium. The mobile phase arriving at the injector
will contain essentially no oxygen, but the solvent in which the sample
is dissolved will contain oxygen unless it has been degassed immediately
prior to injection.
Such artifacts
are particularly noticeable when a UV detector is operated at high sensitivity
and at a low wavelength. They also occur with electrochemical detectors,
where the peaks can be positive or negative.
The following
procedure will show if oxygen is the problem. First remove a 1 mL aliquot
of mobile phase from the chromatograph's reservoir. Immediately inject
a blank using some of this solvent. Observe the peaks. They may be very
small or nonexistent. Now agitate the remaining aliquot for at least
two minutes to equilibrate it with air. Inject another blank using this
solvent. Observe the peaks. If the peaks of the second injection are
larger, and correspond to the retention times of the artifacts, it indicates
that the problem is differences in oxygen concentration.
You can
further confirm this by making blank injections of mobile phase that
contain differing amounts of oxygen. For example, compare injections
of mobile phase that has been sparged (in the vial prior to taking it
into the syringe) with helium, nitrogen, air, or pure oxygen. The relative
amount of oxygen in these three gases is 0.0, 0.2, and 1.0, so the peak
heights in the three cases should have this ratio.
|