Comments
To
remove air from the detector, pump degassed mobile phase and put a back-pressure
regulator at the detector outlet, to produce pressure typically
< 0.7 MPa (6.8 bar, 100 psi) within the cell. Adjust the injector
so it does not inject air (see Symptom #10).
In isocratic
analysis, you can determine if a contaminant is being desorbed from the
top of the column instead of coming from the injector by injecting a blank.
Do this by connecting tubing between the injector and column, a volume
which will make an observable increase in retention time. For example,
100 cm of 0.8 mm (.030 in) OD tubing contains 450 µL, shifting retention
about 0.5 minute at 1 mL/min. If the retention times increase, the substances
are from the injector or sample. If the retention times do not change,
the substances are from the column. Confirm by connecting the tubing between
the column and detector. If the retention times now increase, the substances
are from the column. The tubing will cause the peaks to be broader, but
hopefully the retention times will still be measurable. This test only
works with isocratic analysis.
You can confirm
that peaks in a gradient run are not associated with the injector by connecting
the column directly to the pump (bypassing the injector) and running the
gradient.
|
Non-injector
Artifact Peaks
A blank injection can produce artifact peaks that are not really coming
from the injector.
If the
unknown peak occurs instantaneously with the movement of the injector
handle (i.e., a retention time of zero minutes), then it is probably
due to a pressure disturbance in the detector. This disturbance is caused,
in some cases, by the brief flow interruption that occurs each time
the handle is turned. Whereas most properly functioning detectors will
not respond to such a pressure disturbance, a detector which has become
over-sensitive will respond. Oversensitivity is caused by a gas bubble
in the cell or a cracked window. If you can produce a detector response
by making slight changes in the cell pressure (partially restricting
the outlet), this condition probably exists.
If the
unknown peak has a retention time greater than zero (quite often it
is an unretained peak, i.e., k'=0), then probably a substance is actually
being "injected" onto the column. However, the substance is not necessarily
coming from the injector. It may be coming from contamination of the
column frit or the top of the packing. Such contamination can be desorbed
by an injection in which the sample solvent has a composition different
from the composition of the mobile phase. It may also be that such contamination
can be desorbed by the flow transient caused by handle movement.
Peaks are
sometimes observed in gradient analysis with a blank injection. These
ghost peaks usually come from the initial or final eluent. Using purer
solvents or shortening the equilibration time between runs may eliminate
the problem.
|