Comments
To
clean the injector: Test after each step to see if the problem is solved.
Contamination often is not the problem and decontamination can be time
consuming. You may want to rule out other causes first.
1) Flush
the needle port, vent, and tube #6 using a variety of solvents. 2) Remove
the stator, stator face assembly, and rotor seal (do not disassemble further).
Thoroughly clean their passages using a variety of solvents, preferably
in an ultrasonic cleaner. 3) Replace the rotor seal with a new one of
the same polymer composition. 4) Replace the rotor seal with a new one
of a different polymer composition. In either case, use only genuine
Rheodyne parts.
Although
we have said (previous page) that you should flush
only in INJECT, it is helpful to flush also when in the LOAD position
if you are troubleshooting Cause D. This will clean out vent tube #6,
a possible source of contamination that can diffuse back into the loop.
Peaks obtained
in the confirming experiments of Cause D can sometimes come from the column,
not the injector. See Appendix C.
|
Symptom
#5 (Continued)
Cause
C
Trace contaminants within the injector are loaded. For example, contaminants
in the needle seal (see Fig. 20) are pushed into the loop as the needle
enters the seal. When sample is dispensed from the syringe, the contaminants
will be flushed completely out of the loop only if a considerable excess
of sample is loaded. Contaminants will remain when partial-filling is
used.
Solution
Clean the injector (see comments), however other solutions are sometimes
easier. If you are using partial-filling, try complete-filling and use
a smaller loop if necessary.
![](../images/fig20.gif)
Fig. 20. Droplets
of sample remaining in needle port and needle seal.
Cause
D
Trace contaminants within the injector are "injected" simply by switching
from LOAD to INJECT or from INJECT to LOAD.
To confirm:
Flush in the LOAD position (an exception to the rule, see comments on
previous page) to clean vent tube #6 (see Fig. 21). Then turn to INJECT
and wait for any peaks to elute. Ignore them. Now flush normally. While
still in INJECT, insert a very clean, empty syringe and keep it in place
to prevent remaining contaminants in vent tube #6 or the needle seal
from siphoning or diffusing into the loop. Turn the handle to LOAD and
start the recorder without turning to INJECT. Wait for peaks to elute.
Observe them. Now, with the syringe still in place, turn the handle
to INJECT and start the recorder. Observe the peaks. Peaks in either
or both of these runs indicate internal contamination, most likely on
the rotor seal.
|