Comments
There
are many non-injector causes of pressure changes. They usually occur as
random events, not associated with movement of the injector handle. Examples
are a leaky check valve in the pump or a partially plugged filter in the
mobile phase reservoir.
If the system
pressure increases when the injector is switched, see the Blockage
section.
|
Decreased
System Pressure
This section discusses the causes of a loss of system pressure that
occurs when the injector is switched from LOAD to INJECT or from INJECT
to LOAD. Problems involving an increase in system pressure are discussed
in the Blockage section.
|