Comments
To
flush the needle port: Use about 1 mL of mobile phase. Use a Needle
Port Cleaner (Rheodyne Part Number 7125-054) and only when the handle
is in INJECT. This puts fluid in the entire length of the needle port
guide, tube, and needle seal. The cleaner does this because it
seals against the front of the guide. A fully inserted needle flushes
none of these parts. If you have no cleaner, use a luer tip syringe. See
the illustration in Fig. 18.
In routine
operation, remove the syringe after turning to INJECT, but keep the handle
in this position so the loop is continuously flushed with mobile phase.
To keep the port filled with fluid, flush every ten or twenty injections.
Air in the
loop prior to loading is usually pushed out through vent tube #6 when
using the complete-filling method. But, when using partial-filling, not
enough sample enters to displace the air; Symptom #10
is therefore more likely to occur when using this method.
The decrease
in pressure is instantaneous because the air compresses immediately. It
takes the pump awhile to replace the loss of volume and to return to normal
pressure. Large loops can produce a large loss of pressure which takes
a longer time to return to normal.
|
Symptom
#10
System pressure decreases instantly when the injector is switched from
LOAD to INJECT, and then slowly returns to a normal level:
- If there
is only a small and momentary drop in system pressure, see Cause
A.
- If there
is a large and momentary drop in system pressure, see Cause
B.
Cause
A
Air is present in the needle port, specifically the needle seal (see Fig.
32). When the syringe is inserted, the needle pushes a small air bubble
into the loop. Upon injection, the system pressure collapses the bubble,
causing pressure to drop momentarily.
Solution
Remove the air from the needle port by flushing (see comments). Keep
the needle port filled with fluid by occasional flushing.
Fig. 32. Air
(yellow) present in the needle port is pushed by the syringe during
loading into the sample loop.
|