Comments
To
flush the needle port and fill it: Use about 1 mL of mobile phase. Use
a Needle Port Cleaner
(Rheodyne Part Number 7125-054) attached to a luer tip syringe. Flush
only when the handle is in the INJECT position, so flow goes directly
out vent tube #5 and bypasses the loop. This procedure cleans the entire
length of the port guide, tube, and needle seal-filling it with
liquid. If you have no cleaner, use a luer tip syringe. See the illustration
with Fig. 18.
Mobile phase
and excess sample displaced from the loop during loading exits via vent
tube #6. It can easily become plugged with buffer salt crystals. If this
happens frequently, you should periodically flush the injector in the
LOAD position to prevent build up of crystals.
The syringe
should be left in the needle port until after injection, otherwise some
or all of the sample may siphon out of the loop. See also Symptom
#12 for related problems.
|
Symptom
#9
All
peaks are very small, or no peaks are observed.
While turning
the handle, observe the shaft to which it is attached:
- If the
shaft does not rotate, see Cause A.
- If the
shaft rotates a full 60¡:
In the INJECT position, fill the needle port by flushing (see comments
and Fig. 29). Turn to LOAD and insert the syringe (a little solvent
will appear at the port). Load sample while observing the port and vent
tube #6.
- If solvent
comes out the needle port while loading, but none or very little comes
out vent tube #6, sample is not entering the loop. See Cause
B.
- If solvent
does not come out the needle port, but does come out vent tube #6, sample
is entering the loop. See Cause C.
Fig.
29. Fill the needle port with solvent using the needle port cleaner.
|